Manufacturer of Petroleum Test Equipment since 1873

Product   

 Flash-Point with Closed Cup - Pensky-Martens Method - PMA 4 (automatic)

Standard(s): 

ASTM D 93-A, ASTM D 93-B, ASTM D 93-C, ASTM D 6751, ISO 2719-A, ISO 2719-B, ISO 15267, DIN 51758 (obs.), EN 22719 (obs.), FTM 141-4293, FTM 791-1102, IP 34-A, IP 34-B, JIS K2265-3, NF M07-019 (obs.), AASHTO T73 , AASHTO T172 , BS 684-1.17, BS 2839 (obs.)


 

Product group(s): Flash Point

User group(s): Biodiesel, Biofuel, Bitumen, Engine Oil, Fuel, Jet Fuel, Lube Oil, Mastics, Paint, Power Plant, Regulatory Authorities, Solvent, Turbine Fuel, Turbine Oil, Varnish

Scope:

The flash-point determination of petroleum products is using:

Method A:
for distillate fuels (diesel, kerosene, heating oil, turbine fuels), new lubricating oils, paints, vanishes and other homogeneous liquids.

Method B:
for residual fuel oils, cutback residual, used lubricating oils, non-homogeneous materials like mixtures of petroleum liquids & solids, surface-film building petroleum liquids or liquids with a viscosity above 5.5 cSt at +40°C.

Method C:
for biodiesel (FAME, B100). Electronic flash point detection required.

Used in shipping & safety regulations to define flammable and combustible materials.


  • automatic Barometric Pressure Correction
  • 2 User Definable Test Programs
  • internal memory of 99 test results
  • Milli-test insert with a volume of approx. 15 ml
  • swivel-around Multi-Function-Head - ONE- TWIST