Manufacturer of Petroleum Test Equipment since 1873

Product   

 Dielectric Breakdown Voltage by Electrodes (automatic)

Technical Data

Max. temperature: Operation 0 to +50°C, (storing -20 to +50°C)
Voltage:
True RMS measurement,
Resolution 0.1 kV, Accuracy ± 1 kV
Switch-off current:
Switch-off time:
programmable 1...5 mA
<1ms
Breakdown monitoring: detecting insulating liquids with insulation resistance below 30 MOhms
Voltage withstand tests: Rate of rise, hold voltage and hold time
selectable in 5 steps
Rate of rise: 0,1... 10 kV/s programmable, in automatic mode selected by testing standard
Power consumption: Max. 900 VA
Dimensions: 573 x 450 x 390 mm (W x H x D)