Manufacturer of Petroleum Test Equipment since 1873

Product   

 Flash-Point with Closed Cup - Pensky-Martens Method -
PMA 4 with Sample Changer
(automatic)

Technical Data

Temperature Range:
   - PMA & Sampler:
approx. +40 to +360 °C (+104 to +680 °F)
   - PMA & Chilled Sampler:
below +40 to +360 °C (+104 to +680 °F)
Programs:





(acc. to international standards)
5x ASTM-Standards
4x ISO-Standards
2x Rapid-Heating
2x Search Run
2x User Defined
Ignition Type:
gas and electric (included)
Stirring speeds: 60 to 250 rpm (adjustable)
Sensing System: differential-thermocouple
Barometric Pressure Sensor: automatic correction of the measured values
Safety: overheat protection, automatic shut-off
Hardware Clock: included
Interface:

RS-232 for printer
RS-232 for computer download
(USB-converter available)
Gas Connection:

for propane/butane or
natural gas (max. 0.05 bar)
Display:
°C or °F (selectable), large LCD
Dimension (WxDxH):
   - PMA & Sampler:
  650 x 470 x 650 mm, Weight: 35 kg
   - PMA & Chilled Sampler:
1000 x 470 x 650 mm, Weight: 66 kg
Power Consumption: approx. 900 Watts
Power Supply: 230/115 V, 50/60 Hz (selectable)